FTIR Conversion Monitoring for Coatings: Work with a Validated Band Ratio

Calculate an illustrative FTIR conversion from a reactive-band ratio, validate the reference signal, and separate reaction tracking from product performance.

Updated · NANOLAB Knowledge

FTIR can track a chemical change during curing when the relevant spectral signal is identified and the measurement method is suitable. A declining peak is not automatically a universal percentage-cure reading. The selected chemistry, reference signal and sampling procedure determine what the calculation means.

Choose a signal that answers the chemistry question

Thermo Fisher’s application note AN50752 demonstrates a reactive-band to internal-reference-band ratio for a particular UV-curable ink. The reference band was selected because it did not participate in that ink’s curing reaction. Those peak positions are not a ready-made method for every ceramic coating, silane system or polymer.

Ask the analyst to establish which feature represents the reaction of interest, whether other components overlap it, and whether the proposed reference remains suitable throughout the experiment. ATR and transmission sampling also need different practical controls.

A worked ratio calculation

For a validated signal that decreases with consumption of the reactive group, define R as the selected reactive-band measurement divided by the reference-band measurement. An illustrative conversion calculation is:

Conversion (%) = 100 × [1 − R(t) ÷ R(0)]

Here R(0) is the starting ratio and R(t) is the ratio at the stated elapsed time. Use the same validated peak-height or integrated-area treatment throughout; do not switch between them mid-study.

Hypothetical spectrum Reactive signal Reference signal Ratio
Starting sample 0.80 0.40 2.00
Later sample 0.18 0.30 0.60

The calculated conversion is 100 × (1 − 0.60 ÷ 2.00) = 70%. Looking only at the reactive signal would give a 77.5% decrease. The difference shows why the chosen normalization matters. These fictional absorbance measurements are not NANOLAB laboratory results.

Investigate the changing reference

The numerical example assumes the ratio method has been validated for its sampling conditions. If the reference changes in a real experiment, investigate why: sample contact, effective path, specimen variation or actual chemistry may be involved. A ratio does not repair every problem simply because the spreadsheet returns a percentage.

Define baseline treatment and spectral regions before comparing samples. Preserve raw spectra so another analyst can inspect overlap, saturation, background contributions and changes hidden by automatic rescaling.

Collect a time series with a defensible starting point

  1. Record when mixing, application or exposure begins, and when the first usable spectrum is collected.
  2. Record temperature and any environmental variable relevant to the curing mechanism.
  3. Specify whether the same specimen is followed or separate specimens are measured at each time.
  4. Include repeat preparations so handling variation is visible.
  5. Assess method limitations before interpreting a plateau as a chemical endpoint.

If substantial reaction occurs before the first spectrum, the calculated percentage is relative to that measured starting condition unless a justified correction is used. Do not label it absolute conversion from an unmeasured uncured state.

Connect conversion to the intended coating property

Reaction tracking can support process development, but it does not replace adhesion, appearance or durability testing. A chosen spectral endpoint becomes a useful release criterion only when its relationship to the relevant product requirement is established. See what FTIR can and cannot prove about a coating.

Application reference: Thermo Fisher Scientific, Polymer Workflow and Troubleshooting, AN50752, pages 14–15. The example above illustrates the calculation rather than transferring the ink method to a NANOLAB product.