Private Label & OEM3 min read

From Electrophoretic Mobility to Zeta Potential: Check the Model and Units

Audit a mobility-to-zeta conversion with a worked Henry-equation example. Check model assumptions, permittivity, viscosity and units before comparing reports.

Updated · NANOLAB Knowledge

Electrophoretic mobility and zeta potential are different reported quantities. A laboratory converts mobility into zeta potential using a model and dispersant properties. Before comparing two coating-dispersion reports, check those inputs: changing the conversion can change the reported millivolts without changing the measured mobility.

Write the conversion with explicit units

In the Henry relationship, μe = 2εζf(κa)/(3η), so ζ = 3ημe/(2εf(κa)). Use a consistent SI calculation:

  • μe: electrophoretic mobility, in m²/(V·s).
  • η: dispersant dynamic viscosity, in Pa·s.
  • ε: absolute dispersant permittivity, in F/m; this is ε0εr, not the dimensionless relative permittivity alone.
  • f(κa): dimensionless Henry function, with particle radius a and inverse Debye length κ.
  • ζ: zeta potential, obtained in volts; multiply by 1,000 for millivolts.

The Hückel limiting value is f = 1 for small κa; the Smoluchowski limiting value is f = 1.5 for large κa. These are applicability limits, not settings to select according to which answer looks better. Ask the laboratory to justify its model for the actual sample and medium.

Worked example: unchanged mobility, different conversion

This is hypothetical arithmetic, not a NANOLAB measurement or a recommended test medium. Suppose μe = −2.0 × 10−8 m²/(V·s), η = 1.0 × 10−3 Pa·s and ε = 7.0 × 10−10 F/m.

Assumed conversion Calculation Reported result
f = 1.5 (−6.0 × 10−11)/(2.1 × 10−9) V About −28.6 mV
f = 1 (−6.0 × 10−11)/(1.4 × 10−9) V About −42.9 mV

The magnitude changes by 50% solely because the assumed function changes. Neither row demonstrates improved dispersion stability. Both cannot be treated as interchangeable interpretations of an unspecified sample. Resolve the appropriate model before comparing batches or discussing a specification.

Audit a report before comparing suppliers

  1. Request the mobility result. Keep its sign, units and original precision alongside the converted zeta potential. If only millivolts are supplied, request the underlying result and conversion settings.
  2. Record the tested specimen. Identify the formulation revision, dilution, dispersant, temperature and relevant pH or conductivity information. A changed preparation is a changed comparison.
  3. Check the settings. Record the model, viscosity and permittivity values and their temperature basis. Distinguish instrument fields asking for relative permittivity from an equation requiring absolute permittivity.
  4. Resolve discrepancies in order. First reconcile units and calculation inputs; then examine sample preparation and measurement quality. Do not attribute every difference to chemistry.

For a spreadsheet check, explicitly convert 1 mPa·s to 0.001 Pa·s. A mobility reported as 1 μm·cm/(V·s) equals 1 × 10−8 m²/(V·s). Keep these conversion cells visible instead of burying factors inside the final formula.

Keep the model separate from product approval

Henry conversion is not a universal description of every concentrated or complex formulation. Malvern Panalytical documents additional electrokinetic models; the laboratory must establish which treatment is suitable. This calculation does not prescribe dilution or prove that a selected method preserves the original dispersion.

For the next decision, read why zeta potential cannot independently establish shelf life. If a particle-size report also changes, review viscosity inputs in DLS; that is a separate measurement and calculation.

Primary references