Audit a mobility-to-zeta conversion with a worked Henry-equation example. Check model assumptions, permittivity, viscosity and units before comparing reports.
Updated · NANOLAB Knowledge
Electrophoretic mobility and zeta potential are different reported quantities. A laboratory converts mobility into zeta potential using a model and dispersant properties. Before comparing two coating-dispersion reports, check those inputs: changing the conversion can change the reported millivolts without changing the measured mobility.
In the Henry relationship, μe = 2εζf(κa)/(3η), so ζ = 3ημe/(2εf(κa)). Use a consistent SI calculation:
The Hückel limiting value is f = 1 for small κa; the Smoluchowski limiting value is f = 1.5 for large κa. These are applicability limits, not settings to select according to which answer looks better. Ask the laboratory to justify its model for the actual sample and medium.
This is hypothetical arithmetic, not a NANOLAB measurement or a recommended test medium. Suppose μe = −2.0 × 10−8 m²/(V·s), η = 1.0 × 10−3 Pa·s and ε = 7.0 × 10−10 F/m.
| Assumed conversion | Calculation | Reported result |
|---|---|---|
| f = 1.5 | (−6.0 × 10−11)/(2.1 × 10−9) V | About −28.6 mV |
| f = 1 | (−6.0 × 10−11)/(1.4 × 10−9) V | About −42.9 mV |
The magnitude changes by 50% solely because the assumed function changes. Neither row demonstrates improved dispersion stability. Both cannot be treated as interchangeable interpretations of an unspecified sample. Resolve the appropriate model before comparing batches or discussing a specification.
For a spreadsheet check, explicitly convert 1 mPa·s to 0.001 Pa·s. A mobility reported as 1 μm·cm/(V·s) equals 1 × 10−8 m²/(V·s). Keep these conversion cells visible instead of burying factors inside the final formula.
Henry conversion is not a universal description of every concentrated or complex formulation. Malvern Panalytical documents additional electrokinetic models; the laboratory must establish which treatment is suitable. This calculation does not prescribe dilution or prove that a selected method preserves the original dispersion.