Contact-angle software can fit a drop contour using different mathematical models. Record the selected method so a software-setting change is not mistaken for a coating change.
Updated · NANOLAB Knowledge
KRUSS describes several fitting approaches with different applications. The laboratory should select a suitable method for the observed drop and measurement arrangement, rather than choosing whichever fit produces the preferred angle.
Illustrative example: the same images are reanalysed with another model and yield different angles. The new values require a method explanation; they are not new physical observations of improved coating performance.
Do not compare unexplained software outputs as if they were identical procedures. A qualified reviewer should interpret the fit quality and the limitations of the actual measurement.
Technical references: KRUSS: drop-shape analysis. These sources describe their stated methods or products; they do not certify NANOLAB products or establish compatibility with an unlisted surface.
Ask NANOLAB about the exact product and intended use. Use the current product documentation for formulation-specific instructions.